A procedure to use the RAD7 detector for measuring 222Rn in soil gases exceeding instrumental limits: an application to chemically aggressive fumaroles of the Campi Flegrei area
Main Article Content
Abstract
This report proposes a properdesigned method for the RAD7 detector to measure 222Rn in gases discharged from sulfurous fumaroles, tested at temperatures between 94°C and 162°C.
RAD7 is one of the most widely used commercial instruments for...
RAD7 is one of the most widely used commercial instruments for...
Article Details
How to Cite
Iovine, R. S., Avino, R., Minopoli, C., Cuoco, E., Caliro, S., Galli, G., & Piochi, M. (2023). A procedure to use the RAD7 detector for measuring 222Rn in soil gases exceeding instrumental limits: an application to chemically aggressive fumaroles of the Campi Flegrei area. Rapporti Tecnici INGV, 473. https://doi.org/10.13127/rpt/473